Exiting SmartGuard™ Auto Mode
Device: MiniMed™ 770G insulin pump
The pump will exit SmartGuard™ Auto Mode:
- Auto Mode has been in Safe Basal for 90 minutes.
- A High SG alert has occurred.
- You have not cleared any suspend event messages within four hours.
- You manually turned off the Sensor feature or disconnected the transmitter.
Some alarms cause the pump to exit SmartGuard™ Auto Mode and also turn the Auto Mode feature off. Auto Mode is turned off if an alarm initiates a pump reset. If this occurs, you will no longer see the SmartGuard™ Auto Mode shield on your Home screen. You must turn the Auto Mode feature on again and go through a five-hour warm-up period.
You can turn off Auto Mode at any time.