URGENT MEDICAL DEVICE CORRECTION


Pump Delivery Volume Accuracy (DVA) during Changes in Air Pressure


Pump Delivery Volume Accuracy (DVA) during Changes in Air Pressure - Notification Download

Insulin Pump Model/CFN Number
Paradigm™ MMT-512, MMT-522, MMT-523, MMT-551, MMT-712, MMT-715, MMT-722, MMT-723, MMT-751
MiniMed™ 630G Insulin Pump MMT-1715, MMT-1755
MiniMed™ 670G Insulin Pump MMT-1760, MMT-1780
MiniMed™ 770G Insulin Pump MMT-1880, MMT-1890
MiniMed™ 780G Insulin Pump MMT-1884, MMT-1894

January 2025

Medtronic is contacting you about an “Urgent Medical Device Correction” regarding your MiniMed™ insulin pump. During quality testing performed by Medtronic, we recently found that changes in air pressure can cause unintended insulin delivery. For example, air pressure in an airplane can change rapidly during flight, which may cause expansion of tiny air bubbles inside the insulin reservoir.

Issue Description:

  • When air pressure decreases (e.g., during flight takeoff), more insulin may be released than expected. Additionally, unintended insulin may be released even if the pump’s delivery is suspended or programmed to zero units per hour.
  • When air pressure increases (e.g., during flight landing), less insulin may be released than expected.

Risk to Health:

The changing air pressure conditions could result in more insulin being delivered during flight takeoff, potentially leading to hypoglycemia, or less insulin being delivered during flight landing, potentially leading to hyperglycemia.

Between July 2003 and May 2024, Medtronic has received 138 complaints potentially related with this issue, 19 of which reported serious injuries, but none were confirmed to be related to this issue.

It is important to monitor your glucose frequently while flying and be prepared to treat hypoglycemia or hyperglycemia. Individuals with lower daily insulin doses and those with high insulin sensitivity may experience greater changes in glucose during changes in air pressure than individuals with higher insulin doses and/or lower insulin sensitivity. If you are unsure as to whether this applies to you, it is important that you seek your healthcare professional’s treatment guidance.

Recommended Actions for Pump Users:

  1. Monitor Your Glucose Levels: Check your glucose frequently during activities like air travel, amusement park rides, or other situations where sudden changes or extremes of air pressure, altitude, or gravity may occur.
  2. Discuss how to prepare for situations like this with your healthcare professional. Keep an emergency kit with rapid-acting glucose and backup insulin therapy available at all times.
  3. Respond to Alerts and Symptoms: Pay attention to any alerts from your pump as well as symptoms of hypoglycemia or hyperglycemia. Follow your healthcare professional's treatment instructions in these situations.

Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter by either completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting https://info.medtronicdiabetes.com/pump-delivery-volume to acknowledge this communication.


QR code


Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.


As always, we are here to support you and ensure we are delivering the highest quality products possible. If you have further questions or need assistance, please call the Medtronic 24-Hour Technical Support line at 1-800-646-4633 option 1.

Sincerely,

Julio Salwen
Vice President, Quality
Medtronic Diabetes

Joshua Miller, MD, MPH
Sr. Medical Director, Medical Safety
Medtronic Diabetes

Frequently asked questions


Yes. By FDA definition, this is considered a recall, but it is a correction through a field safety notice letter. The product can continue to be used, and products will not be replaced for this issue.

You received this notification because you were identified as a customer with a MiniMed™ Paradigm™, 600 series, or 700 series insulin pump.

Yes, it is important to acknowledge the safety notice. Please follow the instructions provided in the notice. U.S. patients and all other regions where confirmation is required can click the button in the email or text message stating, "I have read and understand the actions required by me in this Urgent Medical Device Correction" This acknowledgment ensures that you are aware of the critical safety information and allows us to provide you with the necessary support. We will continue to send messages until you acknowledge that you have understand the actions required by you.

Yes, you can continue to use the product with the safety notice following the instructions provided.

No. Your device does not require replacement, please use your insulin pump as normal.

If you need additional troubleshooting assistance, please contact Medtronic 24-Hour Technical Support at 1-800-646-4633.