MiniMed™ insulin pump basal setting programming support
MiniMed™ pump basal setting programming notice
MiniMed™ 630G insulin pump
- MiniMed™ 630G User Guide
- MiniMed™ 630G Transfer Settings Form
- Linking a Transmitter to the MiniMed™ 630G Insulin Pump
- Connecting the Contour®NEXT LINK 2.4 Meter
- MiniMed™ 630G System Support
- Other MiniMed™ 630G Manuals
MiniMed™ 670G insulin pump
- MiniMed™ 670G User Guide
- MiniMed™ 670G Transfer Settings Guide
- Linking a Transmitter to the MiniMed™ 670G Insulin Pump
- Linking a Transmitter to the MiniMed™ 670G Insulin Pump
- Connecting the Contour®NEXT LINK 2.4 Meter
- MiniMed™ 670G System Support
- Other MiniMed™ 670G Manuals
MiniMed™ 770G insulin pump
- MiniMed™ 770G User Guide
- MiniMed™ 770G Transfer Settings Guide
- Linking a Transmitter to the MiniMed™ 770G Insulin Pump
- Linking a Transmitter to the MiniMed™ 770G Insulin Pump
- Connecting the Accu-Chek® Guide Link Meter
- Setup MiniMed™ Mobile App
- Setup CareLink™ Connect App
- MiniMed™ 770G System Support
- Other MiniMed™ 770G Manuals
Frequently asked questions
Yes, by definition, this is a recall for correction. A “recall” does not always mean that you stop using the product or return it to the company. A recall sometimes means that the medical device needs to be checked, adjusted, or fixed, as is the case here.
At Medtronic, customer safety, awareness and customer satisfaction are our top priorities. Medtronic is proactively notifying impacted customers to ensure that they are made aware of the potential safety risks associated with not having saved basal insulin rates programmed into the pump. Medtronic will remind customers of how to ensure these settings are saved in the pump.
No, this notification is not associated with prior urgent recalls issued by Medtronic.
The instructions for how to program basal rates are included in the user guides for the MiniMed™ 600 and MiniMed™ 700 series pumps. However, in April 2021, Medtronic became aware of instances where basal rates were not correctly programmed and/or saved in a new or replacement insulin pump and began researching the issue to determine why this happens and how to help prevent this from happening.
In researching this issue, we identified that it's most likely during the first months of transitioning to a new or replacement pump that the basal rate settings may not be saved in the pump. Therefore, we are proactively reaching out to customers who received new or replacement pumps over the past six months to ensure they have basal rates programmed.
No, your device does not require replacement due to this issue.
All pumps arrive with factory settings and must be programmed with your basal rates and/or other verified settings (i.e., Bolus WizardTM settings, sensor settings, etc.), which must be set up and saved on your pump prior to use.
Check if the basal rate settings are present on your pump. If the basal rate settings are not present on your pump, follow the instructions as indicated in the notice letter and the user guide.